background image

19

a) Press the push-button and pull out the shank/file.

b) When removing the protective silicon cover, pull it straight out 

slowly.

c) When inserting and removing the contra-angle, turn thehandpiece 

power off beforehand.

6.4.3 Cleaning

The cleaning should be performed no later than 24 hours after the 

operation. 

The cleaning can be divided into automated cleaning and manual 

cleaning. Automated cleaning is preferred if conditions permit.

6.4.3.1Automated cleaning

•The cleaner is proved to be valid by CE certification in accordance 

with EN ISO 15883.

▪There should be a flushing connector connected to the inner cavity of 

the product.

▪The cleaning procedure is suitable for the product, and the irrigating 

period is sufficient.

It is recommended to use a washer-disinfector in accordance with EN 

ISO 15883. For the specific procedure, please refer to the automated 

disinfection section in the next section "Disinfection".

Notes:

a) The cleaning agent does not have to be pure water.It can be distilled 

water, deionized water or multi-enzyme. But please ensure that the 

selected cleaning agent is compatible with the product.

b) In washing stage, the water temperature should not exceed 45 °C, 

otherwise the protein will solidify and it would be difficult to remove.

c) After cleaning, the chemical residue should be less than 10mg / L.

6.4.4 Disinfection

Disinfection must be performed no later than 2 hours after the cleaning 

phase. Automated disinfection is preferred if conditions permit. 

6.4.4.1Automated disinfection-Washer-disinfector

·The washer-disinfector is proved to be valid by CE certification in 

accordance with EN ISO 15883.

·Use high temperature disinfection function. The temperature does 

not exceed 134 ° C, and the disinfection under the temperature cannot 

exceed 20 minutes.

·The disinfection cycle is in accordance with the disinfection cycle in 

EN ISO 15883.

Summary of Contents for E-COM

Page 1: ...E COM Endo Motor INSTRUCTION MANUAL...

Page 2: ...ion of product 7 4 Operation instruction 8 5 Oiling of contra angle 14 6 Cleaning Disinfection and Sterilization 14 7 Troubleshooting 23 8 Storage maintenance and transportation 24 9 Environmental pro...

Page 3: ...uring root canal preparation procedure it is used to mold and clean the root canal Features a Wireless handpiece enables more convenient operation b Adopt real time feedback technology and dynamic tor...

Page 4: ...arned not to use household appliances such as electric razors hair dryers etc are not recommended to use this device 1 6 Warnings 1 6 1 Please carefully read this Instruction Manual before first opera...

Page 5: ...n damage to lithium battery and control circuit 1 6 9 The handpiece cannot be autoclaved Use disinfectant of neutral pH value or ethyl alcohol to wipe its surface 1 6 10 Before the contra angle stoppi...

Page 6: ...annot be used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 7 6 Applied part contra angle 1 7 7 The contact duration of applied part 1 to 10 minutes 1 7 8 The te...

Page 7: ...ect contra angle with disinfectant of neutral PH value After disinfection lubricate it with specific cleaning oil Finally sterilize it under high temperature and high pressure 134 2 0bar 2 3bar 0 20MP...

Page 8: ...stallation of file Before starting the device plug the file into the hole of contra angle head While plugging slightly screw the file with one hand and press the push cover of contra angle with anothe...

Page 9: ...s cleaned and disinfected put on a disposable isolation sleeve Take the isolation sleeve out of the isolation sleeve box then insert the isolation sleeve into the motor handpiece from the thin end of...

Page 10: ...consumption d Set speed e Set torque f File system g Operation mode h Torque setting i Real time torque j Set apical stop k File position 4 Operation instruction 4 1 Starting and Stopping a Under the...

Page 11: ...to switchover those memorizing programs And the serial number will change correspondingly 4 3 Speed setting torque setting and operation mode setting Under standby state short press key to enter spee...

Page 12: ...or key to switch operation mode Continuous Rotation Mode Reverse Rotation Mode and Reciprocating Motion Mode are for options There will be a tick indication while setting to Reverse Rotation Mode Long...

Page 13: ...W3 ONE W3 Single Press key to get into the interface of corresponding file system Press key or key to select needed file model the corresponding rotation direction speed and torque value Press the ke...

Page 14: ...r charging please unplug the power adapter Note Please use the original power adapter for charging 4 6 Contra angle calibration setting After replacement of contra angle the contra angle shall be cali...

Page 15: ...wise rotation Load value is lower than preset torque value b Counterclockwise rotation Load value is higher than preset torque value c Clockwise rotation Load value is lower than preset torque value a...

Page 16: ...contra angle for 2 3s till the oil flow out of contra angle head part 3 Vertically place the end part of contra angle or tilt the contra angle to let go the redundant oil under gravity Warnings Handp...

Page 17: ...cable after sterility 6 2 7 The water quality has to be convenient to the local regulations especially for the last rinsing step or with a washer disinfector 6 2 8 To sterilize the endodontic files re...

Page 18: ...fection must be performed within 10min before use b The disinfectant used must be used immediately no foaming is allowed c In addition to 75 alcohol you can use non residue disinfectants such as Oxyte...

Page 19: ...acturer takes no responsibility The products may not be exposed to temperature above 138 Processing limit The products have been designed for a large number of sterilization cycles The materials used...

Page 20: ...ater used here must be pure water distilled water or deionized water 6 4 2 Preparation before cleaning Steps Tools tray soft brush clean and dry soft cloth 1 Remove the shanks files 2 Remove the isola...

Page 21: ...e please refer to the automated disinfection section in the next section Disinfection Notes a The cleaning agent does not have to be pure water It can be distilled water deionized water or multi enzym...

Page 22: ...e should be suitable for the product to be treated The flushing period should be sufficient 5 10 minutes Pre wash for 3 minutes wash for another 5 minutes and rinse it for twice with each rinse lastin...

Page 23: ...he drying temperature should not exceed 138 C c The equipment used should be inspected and maintained regularly 6 4 6 Inspection and maintenance 6 4 6 1 Inspection In this chapter we only check the ap...

Page 24: ...zation procedures are prohibited The steam sterilizer complies with EN13060 or is certified according to EN 285 to comply with EN ISO 17665 The highest sterilization temperature is 138 C The steriliza...

Page 25: ...an atmospheric pressure of 70KPa to 106KPa and a temperature of 20 C to 55 C 2 After sterilization the product should be packaged in a medical sterilization bag or a clean sealing container and store...

Page 26: ...distributor or manufacturer The continuously rotating file is stuck at the root canal Incorrect specification setting Too high load torque of file Choose Reverse Rotation Mode start the handpiece and...

Page 27: ...8 3 3 Avoid solarization and getting wet in rain and snow during transportation 9 Environmental protection Please dispose according to the local laws 10 After service From the date this equipment has...

Page 28: ...erved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner s...

Page 29: ...ZMN SM 069 V1 7 20210427...

Reviews: