the handpiece for a while until the temperature drops.
5 Oiling of contra-angle
Only the original oil injection nozzle can be used for oiling of contra-
angle. After disinfection of contra-angle and before sterilization, oiling
should be conducted under high pressure and high temperature.
1.Firstly, screw the injecting nozzle into jet of oil bottle. (Around 10
circles)
2.Next, plug the nozzle into the end part of contra-angle, and then
grease the contra-angle for 2-3s till the oil flow out of contra-angle head
part.
3.Vertically place the end part of contra-angle or tilt the contra-angle to
let go the redundant oil under gravity.
Warnings
Handpiece cannot be filled with oil.
Cautions
a.To avoid the contra-angle from flying out for the pressure, use hand to
safely hold the contra-angle while greasing.
b.Do not use a swirling nozzle. Swing nozzle can only be used for
injection of gas, not for oiling.
6 Cleaning, Disinfection and Sterilization
6.1 Foreword
For hygiene and sanitary safety purposes,the contra-angle and the
protective silicon cover must be cleaned, disinfected and sterilized