User Manual
WT-UM-08 (v4.4)
11/55
1.4.2
Electrical Specifications
For the WON COSJET ATR S laser system shipped internationally, customers must supply a
suitable plug and receptacle. The power receptacle must be within two meters of the intended
laser system location, and must be grounded. The safety ground wire of the power system (green
or green with a yellow stripe) is an acceptable ground for the laser system, provided that it is
terminated only to an earth ground stake or dedicated ground grid. Poor grounding may interfere
with the operation of the laser system.
WARNING
THE POWER PLUG MUST BE INSTALLED BY A QUALIFIED PERSON, IN
ACCORDANCE WITH IEC REQUIREMENTS AND THE APPROPRIATE
NATIONAL ELECTRICAL CODE.
The input power line should be free of transients (spikes, sags and/or surges). A dedicated branch
circuit is recommended.
Operation of the WON COSJET ATR S on a power line that is not consistently within the
specification may damage the system and may void the warranty.
WON COSJET ATR S Electrical Specifications
Line power requirements:
230VAC, 50Hz
Electric power consumption
4kVA
Foot Switch
After the desired treatment parameters are selected and READY is activated by pushing ready
button, the function of the foot switch, when pressed, allows the Nd: YAG laser to start,
generating laser pulses with the selected properties (fluence/energy, repetition rate etc).
Water resistant:
Ordinary equipment
IPX 0
Foot Switch
IPX 8
Continuous operation with intermittent loading
1.4.3
Environmental Requirements
Before installation of the WON COSJET ATR S laser system, the intended site must be prepared
as described in this section. The site must have sufficient space to accommodate the laser system,
must provide the proper electrical power configuration and receptacles, and must meet the
additional environmental specifications given in the following paragraphs.
Installation of the WON COSJET ATR S laser system is performed by a service representative.
Following installation, a Nurse Consultant instructs designated personnel on the basic operation
and care of the laser. This instruction supplements the more detailed information presented in this
manual. Such instruction is not a substitute for the in-depth clinical training required of a
physician to become proficient in the use of the WON COSJET ATR S laser system.
Sufficient floor space is required for the laser system. Approximately 40 cm of clearance is
required between the rear panel of the laser system and the wall behind it, to allow room for the
power cord and circulation of air from the cooling vents.
Summary of Contents for WON COSJET ATR S
Page 1: ...WON COSJET ATR S User Manual WT UM 08 v4 4 Issue date April 14 2015...
Page 19: ...User Manual WT UM 08 v4 4 19 55 GUI image Articulated arm Spring Type ARM A type...
Page 45: ...User Manual WT UM 08 v4 4 45 55...
Page 48: ...User Manual WT UM 08 v4 4 48 55 Precaution Mark Company name Product Serial Number...