Calibration Unit PGT120 

Part No.: 7100.PGT120.CU 

4 / 8 

2015-05-05 

3

 Footwear test verification (single shoe) - right 

To check the limits of the footwear test connect: 

-  the middle jack of the calibration unit with the 3 mm snap of the PGT

®

 120 

(same symbols 

 ). Use the DK3-socket adaptor which is included. 

-  the  left  jack  of  the  calibration  unit  with  the  red  4  mm  socket  on  the  rear  side  of  the 

PGT

®

 120 

(footwear electrode, same symbols    ). 

Set the marked lever of the rotary switch in succession to the positions mentioned below. 
Press the right contact electrode for each measurement. 

100k Lo 

100k OK 

35M OK 

35M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 

3

.1 

DIP-switch settings for upper limit 70 M

 

Switch 3 

Switch 4 

ON 

ON 

OFF 

OFF 

Set the marked lever of the rotary switch in succession to the positions mentioned below. 
Press the right contact electrode for each measurement. 

70M OK 

70M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 