Calibration Unit PGT120
Part No.: 7100.PGT120.CU
4 / 8
2015-05-05
3
Footwear test verification (single shoe) - right
To check the limits of the footwear test connect:
- the middle jack of the calibration unit with the 3 mm snap of the PGT
®
120
(same symbols
). Use the DK3-socket adaptor which is included.
- the left jack of the calibration unit with the red 4 mm socket on the rear side of the
PGT
®
120
(footwear electrode, same symbols ).
Set the marked lever of the rotary switch in succession to the positions mentioned below.
Press the right contact electrode for each measurement.
100k Lo
100k OK
35M OK
35M Hi
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Display LED
3
.1
DIP-switch settings for upper limit 70 M
Switch 3
Switch 4
ON
5
3
ON
OFF
OFF
8
4
2
1
Set the marked lever of the rotary switch in succession to the positions mentioned below.
Press the right contact electrode for each measurement.
70M OK
70M Hi
○
○
○
○
○
○
○
○
○
○
○
Display LED