Calibration Unit PGT120 

Part No.: 7100.PGT120.CU 

6 / 8 

2015-05-05 

5

 Footwear test in series  

To check the limits of the footwear test connect: 

-

 the middle jack of the calibration unit with the red 4 mm socket on the rear side of the PGT

®

 120 

-

 the left jack of the calibration unit with the yellow 4 mm socket on the rear side of the PGT

®

 120  

5

.1 

DIP-switch setting for upper limit 140 M

 for series 

Switch 8 

ON 

ON 

OFF 

Set the marked lever of the rotary switch in succession to the positions mentioned below. Reset the 
instrument after each measurement by disconnecting the left wire: 

200k Lo 

200k OK 

140M OK 

140M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 

5

.2 

DIP-switch setting for upper limit 70 M

 for series 

Switch 3 

Switch 4 

ON 

OFF 

ON 

OFF 

Set the marked lever of the rotary switch in succession to the positions mentioned below. Reset the 
instrument after each measurement by disconnecting the left wire: 

70M OK 

70M Hi 

 

 

 

 

 

 

 

 

 

 

 

 

Display LED 