Calibration Unit PGT120
Part No.: 7100.PGT120.CU
6 / 8
2015-05-05
5
Footwear test in series
To check the limits of the footwear test connect:
-
the middle jack of the calibration unit with the red 4 mm socket on the rear side of the PGT
®
120
-
the left jack of the calibration unit with the yellow 4 mm socket on the rear side of the PGT
®
120
5
.1
DIP-switch setting for upper limit 140 M
for series
Switch 8
ON
8
5
3
ON
OFF
4
2
1
Set the marked lever of the rotary switch in succession to the positions mentioned below. Reset the
instrument after each measurement by disconnecting the left wire:
200k Lo
200k OK
140M OK
140M Hi
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
○
Display LED
5
.2
DIP-switch setting for upper limit 70 M
for series
Switch 3
Switch 4
ON
8
5
4
OFF
ON
OFF
3
2
1
Set the marked lever of the rotary switch in succession to the positions mentioned below. Reset the
instrument after each measurement by disconnecting the left wire:
70M OK
70M Hi
○
○
○
○
○
○
○
○
○
○
○
Display LED