User’s Manual  – Calibration Unit Safety Pips   

   

Part No. 7100.181.C 

 

 

  _ 

Seite 2 von 2 

 

V

0421

 

 

n

 

Introduction 

The  Calibration  Unit  contains  resistors  and  capacities  within  the  specified  tolerances  for 
testing the Safety Pips. 
The Calibration Unit requires no battery or any external power supply. 

Notice: Remove all connections from the measuring input of the Safety Pip! 

n

 

Checking the Safety Pips 

To check the limits please connect: 

-  Left jack 

of the Calibration Unit with the Safety Pips’s ground. 

 

-  Right  jack  of  the  Calibration  Unit  with  the  Safety  Pips’s  coil  cord  connector  using  the 

banana plug adapter. 

 
Select  sequentially  with  the  marked  leaver  of  the  rotary  switch  each  position  of  the 
Calibration Unit’s left scale (Lo, OK, OK, OK, Hi and Hi). 
The Safety Pips must indicate the results corresponding to the selected position of the rotary 
switch. 

n

 

Checking the Calibration Unit 

Recommended calibration cycle: 3 years 

To check the resistors and capacitances connect a suitable instrument to the sockets of the 
Calibration Unit and select each test position with the rotary switch.  
The corresponding values and tolerances can be taken from the drawing below. 

 

 

 

 

 

 

 

 

 

 

 47pF   

±

10%........................... 

 

 4,7 M

W

  

±

1%............................. 

 

 3,0 M

W

  

±

1%............................. 

 

 2,0 M

W

  

±

1% + 100pF ±10%... 

 

862 k

W

   

±

1%............................. 

 

470 k

W

   

±

1%............................. 

 
 
 
 
 

 

  

Calibration Unit

Safety Pips

Safety Pips

0,4 7M

0,8 62M

2M + 100pF

3,0M

4,7M

47pF

OK

Hi

Hi

OK

OK

Lo