background image

General usage

Always consult your doctor. Self-diagnosis of measurement results 

and self treatment are dangerous People with severe blood flow 

problems, or blood disorders, should consult a doctor before using 

the blood pressure monitor. Cuff inflation can cause internal bleeding. 

Operational factors such as common arrhythmias, ventricular 

premature beats, atrial fibrillation, arterial sclerosis, poor perfusion, 

diabetes, age, pregnancy, pre-eclampsia or renal disease can affect 

the performance of the automated sphygmomanometer and/or 

its blood pressure reading. This device is a precision measuring 

equipment liable to be understood by lay user but it still should be 

handled with care. A long exposure of the device to lint, dust or sunlight 

might reduce its life time or damage it. Damaged cuff or sensor 

might lead to incorrect measurements.
– Do not leave the blood pressure monitor unattended with 

infants or persons who cannot express their consent.

– Do not use the blood pressure monitor for any purpose other than 

measuring blood pressure.

– Do not disassemble the blood pressure monitor.

– Do not operate the blood pressure monitor in a moving vehicle 

(car, airplane).

– Do not use a cellular phone near the deivce. It may result in 

an operational failure.

– Improper continuous pressure of cuff or too frequent measurements 

may interfere blood flow and result harmful injury to the users.

– Do not apply the cuff over the users’ arm bearing a wound or 

medical treatment and so on, as this can cause further influcence 

on the therapy.

AAA alkaline cells usage

– If AAA alkaline cells fluid should get on your skin or clothing, 

immediately rinse with plenty of clean water.

– Use only four AAA alkaline cells with this blood pressure monitor. 

Do not use any other types of AAA alkaline cells.

– Do not insert AAA alkaline cells with their polarities incorrectly aligned.

– Replace old AAA alkaline cells with new ones immediately. 

Replace all four AAA alkaline cells at the same time.

– Do not use new and used AAA alkaline cells together

Replacing the AAA alkaline cells

If the low battery symbol appears in the Withings application, 

replace all four AAA alkaline cells at the same time.

1. Remove the AAA alkaline cells cover at the lower end of 

     the aluminum tube.

2. Install or replace four AAA alkaline cells so that the + (positive) 

     and – (negative) polarities match the polarities indicated on 

     the AAA alkaline cells compartment.

3. Put the AAA alkaline cells cover back in place.
If the device will not be used for a long period of time, it is advised 

to remove the alkaline cells.

CAUTION

Summary of Contents for BP-801

Page 1: ...fore use Keep it for further reference Disclaimer Information in this guide may change without notice The manufacturer assumes no responsibilities for errors that may appear in this guide WIRELESS BLO...

Page 2: ......

Page 3: ...ended use The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches 22cm 42cm Th...

Page 4: ...cellular phone near the deivce It may result in an operational failure Improper continuous pressure of cuff or too frequent measurements may interfere blood flow and result harmful injury to the users...

Page 5: ...ents in water Do not let the device under strong shocks such as dropping the unit on the floor Storage If you are not using the device for an extended period remove the alkaline cells from the alumini...

Page 6: ...r and the collateral device 1 Push the blood pressure monitor button for 3 seconds to turn it off Quit the application and push the blood pressure monitor button to turn it on 2 If the problem occurs...

Page 7: ...o low 1 Check that the Blood Pressure Monitor is correctly positioned on your arm and that mea surement is performed in good conditions 2 Low battery level Unplug the Blood Pressure Monitor Replace th...

Page 8: ...ff position and fasten the cuff correctly ad remeasure the pressure again 2 Check the blood pressure monitor connection to the iOS or Android device The low battery icon is shown in the application Re...

Page 9: ...ferent from that of the clinics or each measurement has various reading 1 The variation is due to the different environments 2 The blood pressure is changing according to the physiological or psycholo...

Page 10: ...285 mmHg Measurement range pulse 40 to 180 beats min Accuracy pressure Within 3 mmHg or 2 of reading Accuracy pulse Within 5 of reading Sensor Semiconductor pressure sensor Operating conditions 10 to...

Page 11: ...ells Accessories AAA Alkaline cells x4 Micro USB to USB adapter Instruction manual Product life 5 years Wireless transmission V4 0 dual mode V2 1 EDR Note Specifications are subject to change without...

Page 12: ...non invasive sphygmomanometers EN300328 V1 8 1 Electromagnetic compatibility and Radio spectrum Matters ERM Wideband transmission systems Data transmission equipment operating in the 2 4 GHz ISM band...

Page 13: ...nce 2013 12 Read this manual before use YA HORNG ELECTRONIC CO LTD ATTEN ELECTRONIC DONGGUAN CO LTD 188 industrial district Ping Shan Administrative District Tang Shia Town Dongguan Guangdong China WE...

Page 14: ...asures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver...

Page 15: ...other equipment and that if adjacent or stacked use is necessary which should be observed to verify normal operation in the configuration in which it will be used Medical Electrical Equipment needs s...

Page 16: ...Monitor system should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RFIEC 61000 4 6 3 Vrms 150 kHz t...

Page 17: ...that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode N A A Mains power quality should be that of a typical commercial or hospital environmen...

Page 18: ...k to the retailer when purchasing a new device Disposing of the product separately prevents possible negative consequences for the environment and for health deriving from inadequate disposal It also...

Page 19: ......