Declaration – electromagnetic emissions and immunity for equipment and systems
that are not life-supporting and are specified for use only in a shielded location
The Wireless Blood Pressure Monitor declaration electromagnetic immunity
The Wireless Blood Pressure Monitor system is intended for use in the electromagnetic environment specified below.
The customer or the user of the Wireless Blood Pressure Monitor system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Conducted
RFIEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
N/A
N/A
Radiated RF IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
N/A
Portable and mobile RF communications equipment should
be used no closer to any part of the EQUIPMENT or SYSTEM
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter. Interference may occur in the vicinity of equipment
marked with the following symbol.