Summary of Contents for BP-801

Page 1: ...fore use Keep it for further reference Disclaimer Information in this guide may change without notice The manufacturer assumes no responsibilities for errors that may appear in this guide WIRELESS BLO...

Page 2: ......

Page 3: ...ended use The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with arm circumference ranging from 9 inches to 17 inches 22cm 42cm Th...

Page 4: ...cellular phone near the deivce It may result in an operational failure Improper continuous pressure of cuff or too frequent measurements may interfere blood flow and result harmful injury to the users...

Page 5: ...ents in water Do not let the device under strong shocks such as dropping the unit on the floor Storage If you are not using the device for an extended period remove the alkaline cells from the alumini...

Page 6: ...r and the collateral device 1 Push the blood pressure monitor button for 3 seconds to turn it off Quit the application and push the blood pressure monitor button to turn it on 2 If the problem occurs...

Page 7: ...o low 1 Check that the Blood Pressure Monitor is correctly positioned on your arm and that mea surement is performed in good conditions 2 Low battery level Unplug the Blood Pressure Monitor Replace th...

Page 8: ...ff position and fasten the cuff correctly ad remeasure the pressure again 2 Check the blood pressure monitor connection to the iOS or Android device The low battery icon is shown in the application Re...

Page 9: ...ferent from that of the clinics or each measurement has various reading 1 The variation is due to the different environments 2 The blood pressure is changing according to the physiological or psycholo...

Page 10: ...285 mmHg Measurement range pulse 40 to 180 beats min Accuracy pressure Within 3 mmHg or 2 of reading Accuracy pulse Within 5 of reading Sensor Semiconductor pressure sensor Operating conditions 10 to...

Page 11: ...ells Accessories AAA Alkaline cells x4 Micro USB to USB adapter Instruction manual Product life 5 years Wireless transmission V4 0 dual mode V2 1 EDR Note Specifications are subject to change without...

Page 12: ...non invasive sphygmomanometers EN300328 V1 8 1 Electromagnetic compatibility and Radio spectrum Matters ERM Wideband transmission systems Data transmission equipment operating in the 2 4 GHz ISM band...

Page 13: ...nce 2013 12 Read this manual before use YA HORNG ELECTRONIC CO LTD ATTEN ELECTRONIC DONGGUAN CO LTD 188 industrial district Ping Shan Administrative District Tang Shia Town Dongguan Guangdong China WE...

Page 14: ...asures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver...

Page 15: ...other equipment and that if adjacent or stacked use is necessary which should be observed to verify normal operation in the configuration in which it will be used Medical Electrical Equipment needs s...

Page 16: ...Monitor system should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Conducted RFIEC 61000 4 6 3 Vrms 150 kHz t...

Page 17: ...that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode N A A Mains power quality should be that of a typical commercial or hospital environmen...

Page 18: ...k to the retailer when purchasing a new device Disposing of the product separately prevents possible negative consequences for the environment and for health deriving from inadequate disposal It also...