10.5. Cleaning and disinfection 

 
High pressure cleaning, machine cleaning, washing tunnel or jet cleaning is prohibited. 
Unplug the mains lead. 
 
Check that all the electrical parts are connected together. All the sockets of the supply box must be used or 
filled, otherwise its watertightness is not guaranteed. 
 
Clean the  electric covers of the jacks and wired control by means of a mop impregnated with a detergent-
disinfectant product. 
 
The medical bed is a non-critical appliance requiring “Low Level” disinfection.  
We draw your attention to the fact that the recommendations below are drawn up according to the rules of 
good practice but are not a protocol. Contact the hospital hygiene department.  
 

 

AIM 

 

To recondition the bed and prevent the transmission of germs from one patient to another.  
To eliminate all organic soiling by:  

- physical action (cleaning)  
- chemical action (disinfection)  

 

 

INDICATION 

 

Physical and bacteriological cleanliness of the bed and its accessories 
 

 

EQUIPMENT 

 

Microfiber wipes 
Detergent or Detergent-Disinfectant (Surface DD with CE mark) and surface Disinfectant (Chlorine 

substances, alcohol base < 30%)  
 

Attention: DD products and Javel water must not be used undiluted. A remanence time must be applied 
according to the disinfectant manufacturer’s instructions (the drying time without human presence is often the 
same as the remanence time) (Refer to the instruction manual)

 

SANIVAP steam appliance with accessories  

 

 

TECHNICAL  

-

  Daily maintenance with a surface DD product in one operation. 

-

  Maintenance when the patient leaves, or periodically, by the process known as 

Bio cleaning

 

observing the 3 operations:  

o

  Cleaning is done by means of a cloth soaked in a surface detergent or Detergent-Disinfectant 

(DD) solution 

o

  Rinsing is done with cloth rinsed in clean water 

o

  Disinfection is done by means of a cloth soaked in a surface disinfectant solution.  

-

  Specific maintenance by specialist contractors after removal of the bed from the establishment:  

o

  Dispose of the packaging after decontamination of the inside by spraying with a Detergent- 

Disinfectant solution  

o

 

Bio cleaning

 operation, or,

 

 

o

  Steam cleaning (accessory with microfiber band) of the flat surfaces and the base slats. 

Change the washing mops regularly to prevent water accumulating. Clean the parts that are 
difficult to access with a steam nozzle (wheels, hinges after opening, corners, etc.). For tubes, 
use the steam nozzle with a microfiber cloth. Never direct the nozzles onto electrical boxes or 
actuators.  

o

  Dry hinges with compressed air  

o

  Attention: Disinfect jacks, electrical boxes and remote controls with a microfiber cloth soaked 

in disinfectant.  

Summary of Contents for AERYS 120

Page 1: ...9_04_01 580046 Anglais 4 Le Pas du Ch teau 85670 SAINT PAUL MONT PENIT TEL 33 0 2 51 98 55 64 FAX 33 0 2 51 98 59 07 Email info medicatlantic fr Site Internet http www winncare fr DUO DIVISYS DO8C P 8...

Page 2: ...6 4 3 NOISE 14 6 5 ELECTRICAL CONNECTION DIAGRAM 15 6 5 1 DOUBLE ELECTRIC STATEMENTS FOLDER DOUBLE STATEMENTS ELECTRIC FOLDING LEGS WITH KNEES 15 6 5 2 DOUBLE ELECTRIC STATEMENTS FOLDER DOUBLE STATEM...

Page 3: ...FTING POLE AND IV STAND 25 10 MAINTENANCE 26 10 1 IDENTIFICATION 26 10 2 INSTRUCTIONS FOR DISMANTLING THE MOTORS 26 10 3 MAINTENANCE 27 10 4 QUALITY INSPECTION OF MEDICAL BEDS 28 10 5 CLEANING AND DIS...

Page 4: ...ANSPORT AND STORAGE For transport the bed should be in its low position on a pallet and strapped and protected The wired control and supply lead should be attached to the bed base The head and footboa...

Page 5: ...e side of the back rest or leg rest if this is not flat In the case of a prolonged more than 50 tilt bust semi sitting position it is recommended to vary the position of the person in bed every 2 hour...

Page 6: ...n the designated room foreseeing an appropriate perimeter of use for the different functions variable height TR etc especially if the bed has a lifting pole or side rails Check that there is sufficien...

Page 7: ...enetration Index 1st number decade Protection against solids 2nd number unit Protection against water intrusion 0 No protection No protection 1 Protected against solid bodies greater than 50 mm Protec...

Page 8: ...ould be at least 30 Electrical fast transients EN 61000 4 4 2 kV for feeders 1 kV for input output lines 2 kV for feeders 1 kV for input output lines The quality of the main power supply must be the s...

Page 9: ...an 3 V m Recommended separation distances between portable and mobile RF communications equipment and the medical bed see references in contents The medical bed see references in contents is intended...

Page 10: ...8 cm minimum 1x Width 88 cm minimum with a high resilience foam of 27Kg m minimum Item Width 120 Width 140 Width 160 MEDIDOM II 2 P319 00 P320 00 P321 00 WINNEA P322 00 MEDIDOM II ELEGANT 2 P324 00 P3...

Page 11: ...00 S A M ACTIV NA A622 00 Angled lifting pole 75 A623 00 Chromed angled lifting pole 75 A631 00 Bed base extension L140 cm NA A632 00 Bed base extension L160 cm NA A637 638 00 EPOXY side rail NA A639...

Page 12: ...the bed is placed on the ground on its wheels It is imperative to raise the kickstand on the ship kit before transporting the bed 6 2 1 Residual risks DANGER RESULT REQUIREMENTS FOR THE USER Trapping...

Page 13: ...l WINNCARE recommends the use of a patient lift or a dining table bases with less than 13 5 cm high Right back rest Bed base Right leg rest Headboard Left back rest Footboard Castors Angled lifting po...

Page 14: ...ot end bed base TOTAL Bed 140 DUO C Version manual crank leg rest without boards 53 kg 31 5 kg 30 kg 114 5 kg Bed 140 DUO P Version electric folding leg rest without boards 57 5 kg 29 kg 38 kg 124 5 k...

Page 15: ...ric folding legs with knees Control box Electrical connection A Remote control or right side if cap infrared option B Remote left or cable side if receiver option IR 1 Right back rest 2 Left electric...

Page 16: ...nuous beeps at the push of a button the remote control It is necessary to reset the electrical assembly This manipulation is only possible with a V0 or V2 type remote control see No of version on the...

Page 17: ...when the bed is empty to familiarise yourself with the bed functions Lifting and lowering of back rest Lifting and lowering of variable height Lifting and lowering of electric folding Simultaneous li...

Page 18: ...t done the patient or another person who leans on the bed may fall 7 ASSEMBLING AND DISMANTLING 7 1 Assembling 1 Brake the wheels 2 Place the foot halt of the base on the base cross brace ensemble to...

Page 19: ...smantling 1 Brake the wheels and place the bed base in low position 2 Unplug from the mains and disconnect the variable height jack and leg rest jack depending on option See stage instructions for dis...

Page 20: ...k rest handle with one hand Push or lift to compensate the pressure exerted by the patient and unhook the clip by the connecting rod with the other hand The back rest jack will then pivot downwards d...

Page 21: ...suscitation 8 2 Leg rest 8 2 1 Leg rest with manual crank C To lift lift the leg rest using the handle at the end To lower relieve the leg rest slightly or to its maximum with the hand to release the...

Page 22: ...To positionning the side rail Insert amount of the side rail folded in the grip jaw on the head side of the bed to hear CLIC Tighten the Rondo screws of the grip jaws on the bed base To remove the sid...

Page 23: ...th both hands until it locks Check that it is properly slot in To lower the side rail Raise the top side rail with one hand Press on the unlocking button with the other hand Support the rail as it low...

Page 24: ...op side rail with both hands until it locks Check that it is properly slot in by trying to get it down without using the unlocking buttons 2 In high position Raise the top side rail with both hands to...

Page 25: ...l Press on the two unlocking buttons with both hands Support the rail as it lowers 9 5 Angled lifting pole and IV stand In case of psycho behavioral problems of the patient it is necessary to fix the...

Page 26: ...uring seals Put the motors back in place and put in the same direction as at the beginning Maximum patient weight Warning refer to user s manual Safe operational load Safety clip No maintenance should...

Page 27: ...ication that the electrical leads are connected all along the metal jambs to prevent shearing of these leads when the variable height is being activated Verification that all of the electrical leads a...

Page 28: ...controls displays lights and locking Bed specific side rails Space between head of bed and siderail 60 mm Space between the half siderails 60 ou 318 mm Condition mounting direction adaptation to bed...

Page 29: ...ed according to the disinfectant manufacturer s instructions the drying time without human presence is often the same as the remanence time Refer to the instruction manual SANIVAP steam appliance with...

Page 30: ...ntenance is 8 years for beds and 5 years for accessories and boards 10 7 Garanties All of our products carry a warranty against any manufacturing defect provided the normal conditions for use and main...

Page 31: ...osition on wheels normal event Press down on the variable height to remove the wheels normal use The selected function of the bed does not work The function is locked on the remote control Unlock the...

Page 32: ...e of the European Union and Council Directive 2006 66 EC of 6 September 2006 on batteries and accumulators and waste batteries and accumulators The complete product and or the easily detachable electr...

Reviews: