ANNUAL MAINTENANCE OF MEDICAL BEDS
IDENTIFICATION
ESTABLISHMENT
VISUAL CONTROL
Not
applicable
Congruent
Improper
FONCTIONAL CONTROL
CPR CONTROL
Not
applicable
Congruent
Improper
ELECTRICAL CONTROL
Not
applicable
Congruent
Improper
Remains in service :
YES
YES subject to the processing of observations
NO
Not
applicable
Congruent
Improper
Obsolescence :
New
Decrepit
*following standard NF EN 60601-2-52
10.4. Quality inspection of medical beds
Identification - label
General cleanliness
Condition and attachment of head and footboards
Lifting pole (fastening, strap, winder, …)
Sleeping surface (bedspring)
Verification of tightenings, diverse nuts and bolts, pins, pivot
State of corrosion, welding
Remote controls (displays, lights and locking)
Bed specific side rails
Space between head of bed and siderail (< 60 mm)*
Space between the half siderails (< 60 ou > 318 mm)*
Condition, mounting direction, adaptation to bed …
Mobile parts (variable height, backrest, legrest, Trendelenburg…)
Remote controls including functions locking
Castors (pivoting, rolling)
Braking
Fixing of siderails and locking in high position (closed)
Absence of sound disturbances (squeaking, lubrications)
Made on :
Name :
Type model / Trademark :
Date of manufacture :
Serial or park number :
Application Environment 1 2 3 4 5
Retracting headboard in case of emergency (App. envir. 1 & 2)
Emergency flattening of backrest
Status of electrical cables, plugs and connectors
Status of electrical equipment (controlbox, actuators,…)
Recurrent test following standard NF EN 62353
(see test report for more details)
COMMENTS
Signature and stamp of technician
Signature of customer