EN
9
3.4. Erasing the scale adjustment
When the moisture content result is
displayed, you can erase the adjustment.
To do that press and hold
F-
button for
about 6 seconds.
3.5. Exceptional result
If the measurement result exceeds the
upper limit of the measurement range, you
will see
HI
on the display and if lower limit
is reached you will see
LO
.
If you get
HI
or
LO
as a measurement
result check if the correct scale is used.
4. Properties of scales
The scales of Wile Coffee - moisture
meter are developed in accordance with
the official methods of definition of
moisture content. To develop the scales
we use the samples which represent
mostly cultivated varieties in standard
growth conditions.
Exceptional growth conditions can affect
the quality and electrical properties of
beans.
Therefore, before the new harvest season
we recommend to check that the readings
of your meter conform to the result of oven
drying. If the measurement result is
different, adjust the result according to the
instructions in paragraph 3.2. This is
especially important when you handle a lot
of beans or when you suspect that the
quality of beans you handle is different