50
Recommended protective distances between portable and mobile RF
communication devices and the AIRcon
The AIRcon Gen2 is intended for operation in an electromagnetic environment in
which RF interferences are controlled� The AIRcon Gen2 customer or the user can
help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communication equipment (transmitters) and the
AIRcon Gen2 - depending on the output power of the communication device, as
indicated below�
Rated capacity of
the transmitter W
Separation distance depending on transmission frequency in m
150KHz to 80MHz
d = 1.2
√
P
80MHz to 800MHz
d = 1.2
√
P
800MHz to 2.5GHz
d = 2.3
√
P
0�01
0�12
0�12
0�23
0�1
0�37
0�37
0�74
1
1�2
1�2
2�3
10
3�7
3�7
7�4
100
12
12
23
For transmitters whose rated capacity is not shown in the above table, the distance
can be determined using the equation for the respective column, where P is
the rated capacity of the transmitter in watts (W) according to the transmitter
manufacturer�
Note 1:
at 80MHz or 800MHz, the separation distance of the higher frequency
range respectively applies
Note 2:
these guidelines may not always apply� The propagation of electromagnetic
waves depends on absorption and reflection by buildings, objects and people�