35
8.14 Water fill level too high
Possible Cause
Measures
Float is defective�
Replace the humidification
chamber�
Error message due to
tilted humidifier�
Place the respiratory humidifier
horizontally or disable water
level detection�
A labeled humidification
chamber is used�
Use non-labeled humidification
chamber or disable water level
detection�
Sensors for water level
detection are dirty�
Remove tube system� Clean
sensors with a clean cloth and
if necessary contact customer
service�
Too much condensate
from tube system
flowed back into the
humidification chamber�
{
{
Empty the tube
system in good time�
{
{
Reduce excess water
in the humidification
chamber manually�
{
{
Disable water level
detection temporarily�
8.15 HW alarm
Possible Cause
Measures
Remove tube system and report
to customer service�