46
Guidance and manufacturer's declaration –
electromagnetic immunity to interference
The AIRcon Gen2 is intended for operation in an environment as specified below�
The customer or user of AIRcon Gen2 should ensure that it is operated in such an
environment�
Radiation
resistance
IEC 60601 –
test level
Conformity
levels
Electromagnetic
environment – Guidelines
Conducted
RF interfer-
ence c�f� IEC
61000-4-6
Conducted
RF interfer-
ence c�f� IEC
61000-4-3
Radiated RF
disturbances
in the imme-
diate vicinity
of wireless
commu-
nication
devices in
accordance
with IEC
61000-4-3
3V
eff
150 kHz to
80 MHz
6V
eff
in ISM
and amateur
radio fre-
quency bands
between
0�15MHz
und 80MHz
10V/m
80 MHz to
2,7 GHz
Included in
table 1
3V
6V
10V/m
80 MHz to
2,7 GHz
Included in
table 1
Portable and mobile radio equipment
should not be operated in closer
proximity to AIRcon Gen2 including
power cord than the recommended
safety distance, which is calculated
according to the equation appropri-
ate to the transmission frequency�
Recommended separation distance:
d = 3,5/3
√
P
d = 3,5/3
√
P
80 MHz to 800 MHz
d = 7/3
√
P
800 MHz to 2,5 GHz
with P as the rated power of the
transmitter in watts (W) according to
the transmitter manufacturer and d as
the recommended separation distance
in meters (m)�
The field strength of stationary radio
transmitters, based on an on-site
examination
a
, should be less than the
compliance level
b
at all frequencies�
In the vicinity of equipment marked
with the following symbol, interference
is possible�
Summary of Contents for AIRcon Gen2
Page 51: ...51 Notes...