49
Test
frequency:
MHZ
Frequency
band
a
MHz
Radio
service
a
Modu-
lation
b
Maximum
power W
Distance
m
Inter-
ference
immunity
test level
2450
2400 up
to 2570
Bluetooth,
WLAN
802�11
b/g/n, RFID
2450, LTE
band 7
Pulse
modula-
tion
b
217Hz
2
0�3
28
5240
5100 up
to 5800
WLAN
802�11 a/n
Pulse
modula-
tion
b
217Hz
0�2
0�3
9
5500
5785
NOTE:
If necessary, the distance between the transmitting antenna and the ME
device or ME system can be reduced to 1 m in order to reach the noise immunity
test level� The 1-meter test distance is permitted according to IEC 61000-4-3�
a
For some radio services, only the frequencies for the radio link from the mobile
communication device to the base station (uplink) were included in the table�
b
The carrier must be modulated with a square wave with 50% duty cycle�
c
Alternatively to frequency modulation (FM), a pulse modulation with 50% duty
cycle with 18Hz can be used, since this, even if not the actual modulation,
would thus represent the worst case�