16
Operating Instructions - CPA2501
Air Data Test Indicator
CPA2501
4.2 Measurement & General Specifications
Basic Instrument
Instrument
Instrument version
Standard: Table top with tilt feet
Option:
-19” rack-mounting for single instrument mount.
-19” rack-mounting for dual instrument mount.
Dimensions
See technical drawing
Weight
12.5 lbs./ 5.7 kg (with all internal options)
Warm-up time
Approximately 15 minutes
Display
Screen
7” color LCD
Resolution
Selectable from 4 to 6 digits, depending on range and units
Data entry
Touch screen keypad
Pressure Measurement Units
psi, psf, osi, atm, inH20@4C, inH20@20C, inH20@60F, mbar, bar, Dy/cm2, pascal,
hPa, kPa, MPa, inHg@0C, inHg@60F, mTorr, Torr, mmHg@0C, cmHg@0C, mHg@0C,
mmH20@4C, cmH20@4C, mH20@4C, mmH20@20C, cmH20@20C, mH20@20C,
mSW, ftH20@4C, ftH20@20C, ftH20@60F, inSW, ftSW, tsi, tsf, g/cm
2
, kg/cm
2
, kg/m
2
, %
of Range, + plus 2 user defined units (multiplier from psi, bar or pascal)
Rate Units
/sec., /min., /hr., /3-hr
Languages
English, German, Spanish, French, Italian, Portuguese, Polish, Russian, Chinese,
Japanese, Korean
Measurement filters
Off, Low, Normal (default), High
Connections
Number of integrated transducer
(selectable)
Standard: 1 reference transducer
Optional: 2nd reference transducers, external transducer, internal barometric reference
Pressure connections
Ranges ≤ 6015 psi: 7/16 - 20 female SAE/MS (adapters provided)
Pressure adaptors
Standard: 1/8 in. FNPT
Optional: 1/4 inch tube, 6 mm tube, 1/4 in FNPT, 1/8 in. FBSP.
Overpressure limits
110 % FS typical, optional external relief valves are available
Voltage supply
Power input requirements
100-120 or 200-240 VAC, 50-60Hz, 24VA max
Switching power supply
Output: 12 VDC, 1.7 A (includes 4 region specific plugs adapters)
Permissible ambient conditions
Storage temperature range
0 to 70 deg C
Operating environment
0 ... 95 % RH (relative humidity, non-condensing)
Operating temperature range
15 ... 45 deg C
Operating Altitude
<3048 meters (10,000 ft)
Communications
Remote interface
IEEE 488, RS-232, USB and Ethernet
Command sets
Mensor, WIKA SCPI
CE conformity and certificates
CE compliance
EN61326-1:2013 electromagnetic compliance
EN61010-1:2010 safety/CB scheme
EN50581:2012 hazardous substances
Calibration
Calibration certificate per ISO/IEC 17025:2005. Accreditation is by the American
Association for Laboratory Accreditation (A2LA).