8. Faults
Faults
Causes
Measures
No output signal
No power supply or cable
break
Check voltage supply and cables,
replace any defective components
found
Electronics failure due to too
high power supply or through
external voltage
Return the instrument to the
manufacturer for repair
Too high, constant
output signal upon
change in temperature
Electronics failure due to too
high power supply or through
external voltage
Return the instrument to the
manufacturer for repair
Signal span too small
Power supply too low
Rectify the power supply
Load too high
Stay within the max. permissible load
Zero point maladjusted
Return the instrument to the
manufacturer for repair
Zero point too low/
too high
Zero point maladjusted
Return the instrument to the
manufacturer for repair
Output signal instable
Power supply too low
Rectify the power supply
CAUTION!
If faults cannot be eliminated by means of the measures listed above, shut
down the instrument immediately, and ensure that pressure and/or signal
are no longer present, and secure the instrument from being put back into
operation inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
9.2 "Return".
8. Faults
Summary of Contents for intelliTHERM TGT70
Page 16: ...16 WIKA operating instructions model TGT70 EN 14047194 03 03 2022 EN DE FR ES...
Page 30: ...14047194 03 03 2022 EN DE FR ES 30 WIKA Betriebsanleitung Typ TGT70 DE...
Page 44: ...44 WIKA mode d emploi type TGT70 FR 14047194 03 03 2022 EN DE FR ES...
Page 58: ...58 WIKA operating instructions model TGT70 14047194 03 03 2022 EN DE FR ES...
Page 59: ...WIKA operating instructions model TGT70 59 14047194 03 03 2022 EN DE FR ES...