Appendix A: Diagnostic Codes
Measurement Status / TransientErr
Diagnostic Status Bits
Set
both
AB/BA
Affects WARNING ERROR
Hex Value
Decimal
Value
Enum / Short Description
Long Description
MeasStatus
Values set f
or Meter
Values
0x00000000 0
No_error
No Error
0x00000001 1
Warning
Measurement Warning
X
0x00000002 2
Error
Measurement Error
X
0x00000004 4
CalSwitch_unlock
Calibration lock open
X
0x00000008 8
System_error
Inconsistency in system
parameters
X
0x00000010 16
FPGA_error
Inconsistency in FPGA
parameters
X
0x00000020 32
Flash_mem_error
Error accessing internal flash
memory
X
0x00000040 64
RTCBattFail
RTC clock battery failure
X
0x00000080 128
Ispare
Not enough power
X
0x00000100 256
Firmware_error
Firmware checksum error
X
0x00000200 512
Factory_defaults
Started from factory defaults
X
0x00000400 1024
ParamCRC_Error
Parameter checksum error
X
0x00000800 2048
Optionboard_timeout
Option board does not respond
X
0x00001000 4096
TestMode
Meter is in test mode
X
0x00002000 8192
DigOutOverflow
LF pulse output overflow
X
0x00004000 16384
TotalStorageInitError
Totalizer storage error - no valid
flash record found on startup
X
0x00008000 32768
TotalPersistError
Totalizer persistence error -
totalizer storage verification failed
X
0x00010000 65536
PTError
PTZ module error
X
PTZStatus
Diagnostic Status Bits
Set both
AB/BA
Affects WARNING
ERROR
Hex Value
Decimal
Value
Enum / Short Description
Long Description
PTZ
Status
0x00000000 0
PTStatusGood
PTZ No errors
0x00000001 1
PTStatusWarning
PTZ Warning
X
0x00000002 2
PTStatusError
PTZ Error
X
0x00000004 4
Spare
Spare
0x00000008 8
Spare
Spare
0x00000010 16
TemperatureOutofRangeError Temperature sensor error or value
out of range
X
0x00000020 32
PressureOutofRangeError
Pressure sensor error or value out
of range
X
0x00000040 64
PTReadTimeOutError
No response from PT option
board
X
0x00000080 128
PTZInputsOutofRangeError
PTZ inputs out of range
X