7.7 Select the overrun and underrun behaviour
Overrun is indicated with 4 bars above.
Underrun is indicated with 4 bars below.
The exception to this is the 4 ... 20 mA input signal, where a measured value
smaller than 1 mA is already indicated as an underrun. This signalises a sensor
failure.
Parameter
Description
no
If the indication range is left, the display will remain on the smallest value
(dI.Lo) or the highest value (dI.HI).
AdC
With the display range being exceeded above or below the limits (dI.Lo/dI.HI),
over- or underrun will be displayed.
rAnG
With the measuring range being exceeded above or below the limits (EnD/
OFFS), over- or underrun will be displayed. The indication range and trans-
ducer range are monitored additionally.
5Pr
The input signal is monitored to ±5 % of the measuring range. The indication
range is also monitored.
10Pr
The input signal is monitored to ±10 % of the measuring range. The indication
range is also monitored.
TYPE
OVEr
[
▲
] [
▼
]
[
☼
]
[
☼
]
no
AdC
rAnG
[
▲
]
[
▼
]
5Pr
10Pr
7. Voltage or current measurement