37
WIKA operating instructions, model CTH6200
EN
2146872.02 07/2021 EN/DE
10. Specifications
10. Specifications
10.1 Digital indicator model CTH6200
Indicator model CTH6200
Electrical connection for temperature probe
Measuring input
1 input
Sensor compatibility
Compatible with temperature probe model CTP6210 and
CTP6290
Connection at CTH6200
4-pin, shielded mini DIN female connector
Digital display
Measuring range
Depending on the connected temperature probe
■
-199.99 ... +199.99 °C [-199.99 ... +199.99 °F] or
■
-199.9 ... +850.0 °C [-328.0 ... +1562.0 °F]
Indication range
Depending on the set resolution of the instrument
■
-199.99 ... 199.99 °C [-199.99 ... 199.99 °F] or
■
-199.9 ... 999.9 °C [-199.9 ... 999.9 °F]
Display resolution
0.01 °C [0.01 °F]
0.1 °C [0.1 °F]
Indication accuracy
≤ 0.03 °C [0.06 °F] for resolution 0.01°
≤ 0.1 °C [0.2 °F] for resolution 0.1°
Type of indication
LC display, for display of values and additional information
Number of lines, digits
2x 4½-digit
Character size
12.4 mm or 7 mm [0.49 in or 0.28 in]
Units
Adjustable between °C or °F
Rated temperature
25 °C [77 °F]
Temperature drift
≤ 0.002 °C/K
Measurement
Measurement type
4-wire-measurement with thermoelectric voltage compensation
Measuring current
0.3 mA
Functions
Measuring rate
2 measurements/s
Mean value filter
1 ... 30 seconds; can be set via menu
Real-time clock
For data logger; can be set via menu
Hold
Holding the last measured value; can be accessed via function
button
Tare
Button only active within the menu