3. Specifications
Input and output
signals
Measuring
range
Absolute measurement
uncertainty (all errors
incl.)
Resistance thermometer
1)
Pt100 (385)
-200.0 ... +800.0 °C
0.33 °C
Pt100 (3926)
-200.0 ... +630.0 °C
0.3 °C
Pt100 (3916)
-200.0 ... +630.0 °C
0.3 °C
Pt200 (385)
-200.0 ... +630.0 °C
0.8 °C
Pt500 (385)
-200.0 ... +500.0 °C
500.0 ... 630.0 °C
0.3 °C
0.4 °C
Pt1000 (385)
-200.0 ... +100.0 °C
100.0 ... 680.0 °C
0.2 °C
0.3 °C
Ni120 (672)
-80.0 ... +260.0 °C
0.2 °C
Cu 10
-100.0 ... +250.0 °C
2.2 °C
Cu 50
-180.0 ... +200.0 °C
0.5 °C
Cu 100
-180.0 ... +200.0 °C
0.3 °C
YSI 400
15.0 ... 50.0 °C
0.1 °C
Resistance
Excitation current
Output
5.0 ... 400.0 Ω
5.0 ... 400.0 Ω
401 ... 1,500 Ω
1,501 ... 3,200 Ω
0.15 Ω
0.1 Ω
0.5 Ω
1.0 Ω
0.1 ... 0.5 mA
0.5 ... 3 mA
0.05 ... 0.8 mA
0.05 ... 0.4 mA
Input
0.00 ... 400.00 Ω
401.0 ... 1,500.0 Ω
1,501.0 ... 3,200.0 Ω
0.1 Ω
0.5 Ω
1.0 Ω
1) Absolute measurement uncertainty based on a 4-wire circuit
For further specifications see WIKA data sheet CT 82.01 and the order documentation.