background image

AquaSpense™ 

and AquaSpense SL™ 

User’s Manual 

Whip Mix Corporation

 

361 Farmington Ave. • P.O. Box 17183  • Louisville, KY 40217-0183 USA

502-637-1451 • 800-626-5651 • Fax 502-634-4512 • www.whipmix.com

Summary of Contents for AquaSpense

Page 1: ...AquaSpense and AquaSpense SL User s Manual Whip Mix Corporation 361 Farmington Ave P O Box 17183 Louisville KY 40217 0183 USA 502 637 1451 800 626 5651 Fax 502 634 4512 www whipmix com ...

Page 2: ......

Page 3: ...s Cap and Bottle Hose 09556 1 Liter Bottle 09555 Special Liquid Cap Assembly includes Cap Special Liquid Hose and Bottle Hose 09552 TABLE OF CONTENTS Front Rear View 2 3 Set Up 4 Programming the AquaSpense SL 5 Programming the AquaSpense Water only 6 Running a Program 6 Special Functions Calibration 7 Scale Mode 8 Set Language 8 Erase a Program 8 Purge Liquid SL Units only 8 Prime Water Hose 9 Mai...

Page 4: ...SPECIAL LIQUID DISTILLED WATER SCALE ARM TO ACCESS HOSE FRONT VIEW 2 Note Water only units do not have the Arm to Access Hose or the Special Liquid bottle or holder ...

Page 5: ...SCALE SOCKET SPECIAL LIQUID DISTILLED WATER POWER CORD SOCKET ON OFF SWITCH REAR VIEW 3 ...

Page 6: ...d extension hose and inline hose connector to allow the use of a larger container such as a one gallon jug Insert the connector end into the clear tubing from the unit and place the hose directly into a container SL Units only The special liquid bottle can be filled with investment liquid or die stone additive Fill the special liquid bottle with liquid secure cap by holding the cap and turning the...

Page 7: ...the numbers or letters to name the program Press the button to advance to the next position Press the button to move back to the previous space 7 Once the name is complete press the Enter button to move to the next screen 8 Use the or buttons to enter the amount of powder in grams 9 Use the or buttons to indicate the amount mL of Special Liquid to dispense per the given powder weight above 10 Pres...

Page 8: ...ess the Enter button 6 Use the or buttons to select the numbers or letters to name the program Press the button to advance to the next position Press the button to move back to the previous space 7 Once the name is complete press the Enter button to move to the next screen 8 Use the or buttons to enter the amount of powder in grams 9 Press the Enter button 10 Use the or buttons to indicate the amo...

Page 9: ...investment or stone SPECIAL FUNCTIONS MENU Calibration The scale arrives calibrated from the factory The calibration information is stored in the scale so changing scales will not necessitate a recalibration The scale should rarely if ever need to be calibrated 1 Obtain a 500 g to 1200 g known weight If needed weigh an object on a separate scale in order to calibrate the dispenser 2 Be sure the ru...

Page 10: ...le 7 Check the weight displayed against the known weight of the object 8 If the weight displayed does not match the known weight then repeat the calibration process Otherwise press Menu or Abort to return to the main menu Set Language 1 Use the or buttons to move the cursor to the desired language The unit can display in English French German Italian and Spanish 2 Press the enter Key 3 The unit wi...

Page 11: ...r container 1 Press the Menu button 2 Use the or buttons to select Special Functions 3 Press the Enter button 4 Use the or button to select Prime Water Hose 5 With a container on the scale press the Start button 6 After air is removed from hose press Abort to end MAINTENANCE Note Some special liquids crystallize when exposed to air and may impede dispensing of special liquid Daily maintenance is r...

Page 12: ...ecial liquid hose into the round opening on the inside left of the arm Allow the hose to extend approx 1 inch below the opening Fig 3 Place your thumb on the hose nearest the white pinch valve and hold with direct pressure Fig 4 Stretch the hose and slide into the pinch valve The hose should slide in easily if not stretch the hose further Thread the other end of the hose through the opening in the...

Page 13: ...nt of water special liquid in the bowl this will not affect accuracy Check alignment of scale FOR SL UNITS ONLY Excessive clicking OR will not dispense special liquid OR Out of Special Liq error Ensure there is sufficient liquid in the bottle Ensure the bottle cap is free of debris and crystallized particles Tightly secure the bottle cap Gently pinch the end of the special liquid hose to free crys...

Page 14: ...nes x 20 characters Operational Altitude Sea level 2 000 meters 6 560 feet Operation Temperature 13º C 41º C 55º F 105º F Maximum Relative Humidity 80 for temperatures to 35º C 95º F decreasing linearly to 50 relative humidity at 35º C 104º F Pollution Tolerance Level 2 in accordance with IEC 664 Recommended Environment Vertical position indoors within operational parameters Warranty 3 years parts...

Page 15: ......

Page 16: ...MPL5600 11 06 ...