27
16.3 Fuse Replacement
DANGER! BE CERTAIN THE UNIT IS DISCONNECTED FROM THE AC POWER
SOURCE.
1. Disconnect the cord-set from the AC power source.
2. Remove the fuse drawer from the unit.
3. Replace with new fuses:
108 - 120 vac unit use (2) 5x20mm, 250V, 2.0 AT fuses,
be sure the
correct cord set is used.
220-240 vac unit use (2) 5x20 mm 240V 1.0 AT fuses
, be sure the
correct cord set is used.