SECG 4.0 | User Manual
(use WhaleTeq CMRR3.0
for these tests)
IEC 60601-2-47:2012
(Ambulatory)
201.12.4, all tests except
CMRR (use WhaleTeq
CMRR 3.0 for these tests)
For all 201.12.1.101
database tests, use
WhaleTeq MECG 2.0 for
these tests.
No known limitations
ANSI/AAMI EC 13
2002/(R)2007/C2008
All performance tests
except CMRR and as noted
left.
See below for Clauses
5.1.4 n) and 5.2.9.1 f), g)
1
General limitation: this equipment is designed for use with isolated ECG circuits, as
are generally provided for medical ECG. If applied to a non-isolated circuit, the noise
may be excessive.
1
5.1.4 n) Fast QRS: The sampling rate is limited to 0.2ms, some distortion of the pulse is possible
below 6ms.
5. 2.9.1 f), g): Functionality is not included at this time; note the test is not applicable to most ECG
systems