![WHALETEQ SECG 4.0 User Manual Download Page 42](http://html.mh-extra.com/html/whaleteq/secg-4-0/secg-4-0_user-manual_967658042.webp)
42
Single Channel ECG Test System | User Manual
4
Software Development Kit
WhaleTeq provides SECG4.0 software development kit. All operating parameters
and options have corresponding commands in the software development kit. The
software development kit contains DLL (Dynamic-link library), which will provide
highly efficient program binding and version upgrade, supports C/C++ header and
C# interface, and can also be integrated with third-party tools and script languages.
5
Calibration, software validation
The WhaleTeq SECG 4.0 has undergone a detailed system validation including
software. A report for this can be provided on request.
Prior to shipping, each unit is tested for component values and output voltages,
using a calibrated precision multi-meter. As WhaleTeq cannot provide ISO 17025
accredited calibration, laboratories which are required to follow ISO 17025 should
perform calibration either periodically or on a before use basis, following normal
procedures and practice. The extent of calibration may be limited depending on the
needs of the laboratory.
As the calibration procedure is complicated, a software assisted calibration mode is
provided. The software sets up the SECG as required for the particular tests, and
instructs the user on what measurement to make (e.g. measure resistance between
RA and RL).