Single Channel ECG Test System | User Manual
2.6.10 Load ECG File
This function is implemented on the
“
Special Waveforms
”
form. In that form, a new button is created and a text box.
The button name in code is
“
Load ECG
”
The
“Load ECG” function supports two formats –
Text and Binary files
Text (*.txt)
- Ascii file, Windows line breaks (LF, CF)
- first line is sample rate (Hz)
- second line number of samples
- following lines are samples in microvolts
(one sample per line)
Binary files (*.bin)
- Bytes 1-2 are sample rate (Hz)
- Bytes 3-6 are number of samples
- Following bytes are samples, 2 bytes per sample
- all data is bigendian (high byte first), 2's compliment
The maximum size of loaded file is 100 million samples.