Westlab 665-090 Product Manual Download Page 7

7

PRODUCT MANUAL

1800 358 101

[email protected]

www.westlab.com.au

+64 9 553 3677

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

Number

Name

Description

1

Display

The display shows the set speed and the time value.  

It also displays the selected program when the unit is 

in Prog. Mode.

2

Speed

Press the ‘Speed’ button to select the speed value and 

use the increment and decrement button to set the 

required speed.

3

Time

Press the ‘Time’ button to select the time value and 

use the increment and decrement button to set the 

required run time.

4/5

Increment/ 

Decrement

Press the ‘Increment/Decrement’ button to set the 

required value of speed, time or program [Prog. Mode].

6

Start/Stop

Press the ‘Start/Stop’ button to start the operation or 

to stop the ongoing operation.

7

Pulse and LED

Press the ‘Pulse’ button once to start the pulse 

operation and press it again to stop the operation.  

The Pulse LED will glow to indicate Pulse is activated. 

It can also be used for programmable pulse mode.

8

Prog

Hold down the ‘Prog.’ button to enter in the program 

mode and use the other required buttons to set the 

parameters for the selected program.

Powering the device ON

1.  Connect the power adaptor to the back of the device
2.  Switch the main power supply to ON
3.  Switch the device ON from the back

ATTENTION: Ensure that the tubes are loaded correctly before use. DO NOT start the 

device if the hinge is open.

Setting the speed

1.  Press the ‘Speed’ button to select the speed value
2.  Press the ‘Increment/Decrement’ button to set the required speed

The speed for operation ranges from 10 RPM to 80 RPM with incremental or 

decremental steps of 1 RPM. The speed values are saved after blinking five times.

Summary of Contents for 665-090

Page 1: ...1 PRODUCT MANUAL Product Manual Roller 6 Tubes 80RPM Max Code 665 090...

Page 2: ...d suspensions 1 3 Features Gentle rocking and rolling system for optimal mixing Adjustable speed ranging from 10 to 80 RPM Timer ranging from 1 minute to 99 mins and infinite mode Maintenance free bru...

Page 3: ...Specification Motor Type Brushless DC motor Number of Tube Rollers 6 Rollers Rocking Motion Rolling and rocking Speed Range 10 to 80 RPM Speed Accuracy 5 RPM Speed Steps 1 RPM Tilt Angle 3 Timer Range...

Page 4: ...o mix explosive or highly reactive substances Do not place the potential hazardous material within the clearance area or envelope Wear your personal protective equipment in accordance with the hazards...

Page 5: ...our feet stand on the surface firmly Avoid installing on a slippery surface or surface prone to vibration 2 Ideal ambient temperature is 20 C 5 C avoid placing the device in direct sunlight 3 Keep the...

Page 6: ...com www westlab com 1 877 822 1455 sales ca westlab com www westlab com Australia New Zealand United States Canada Installing the rollers Follow the exact reverse process to insert the roller back to...

Page 7: ...rt Stop Press the Start Stop button to start the operation or to stop the ongoing operation 7 Pulse and LED Press the Pulse button once to start the pulse operation and press it again to stop the oper...

Page 8: ...ery 30 seconds The default cycle time set for pulse mode is 30 seconds 1 Press the Start Stop button again to stop the ongoing operation at any time The speed will gradually decrease from the set RPM...

Page 9: ...t Stop button to stop the ongoing pulse mode operation Program mode The program mode feature allows the user to save up to nine programs with different speed and time parameters Users can use any of t...

Page 10: ...ved Press the Start button to start the program mode operation Once the set RPM is reached the program Px with x 1 to 9 and its Speed and Time values will display alternatively 7 To stop the operation...

Page 11: ...eed time and other values set by the user When the power restarts the roller will run for the remaining time left with a cycle time of 5 minutes The unit will save the run time every five minutes of o...

Page 12: ...Westlab in every discovery 12 PRODUCT MANUAL...