8. Maintenance 

FAR eye showers must undergo a function test at least once a month. This is set down in 

regulation BGI/GUV-I 850-0 of the BG Chemie. To do this, activate the shower for approx.  

5 seconds. This ensures not only that correct function is tested but that any standing water  

in the shower is replaced by fresh water (protection against microbial contamination).

As part of this function test, the shower should be checked visually for signs of wear and tear 

and any possible leaks and dirt such as e.g. dust or scale deposits. Inspect the hose for any 

twisted, bent or otherwise damaged places. Faulty parts must be replaced immediately.

9. Procedure for leaking hand-held eye showers

After connecting to the cold water system, the hand-held eye shower may have started to 

leak. This is generally due to dirt from the water supply which collects in front of the valve.  

If this has happened, disconnect the handheld eye shower from the supply. Operate the 

shower a few times without pressure and rinse the pipe thoroughly before reconnecting. If the 

fault cannot be remedied by this means the enclosed mesh filter must be fitted in front of the 

stainless steel hose.

12

PRODUCT MANUAL

ORDERING WITH 

WESTLAB IS EASY

Online

www.westlab.com.au

Order online with our easy 

Express Order form for fast, 

secure ordering.

Phone

1800 385 101

Speak to one of our friendly, 

efficient staff from

8.00am - 5.00pm (Mon/Fri)

Email

[email protected]

24 hours, 7 days a week 

convenience.

1800 358 101

[email protected]

www.westlab.com.au

09 972 7682

[email protected]

www.westlab.co.nz

646 568 5391

[email protected]

www.westlab.com

1 877 822 1455

[email protected]

www.westlab.com

Australia

New Zealand

United States

Canada

Summary of Contents for 664-812

Page 1: ...1 PRODUCT MANUAL Product Manual Floor Mounted Emergency Shower Head Eye Wash Code 664 812...

Page 2: ...Coated Surface Protection Hand Foot Operated 100 Italian Made 2 Technical Specifications Application Floor Fixed Test Pressure 9 bars air Working Pressure Max 5 bars Flow Rate 30L min at 1 bar Surfac...

Page 3: ...uidelines 2 Operating pressure The FAR safety showers are designed for an operating pressure of a maximum of 10 bar PN 10 The flow rate is dependent on the available water pressure as well as the cros...

Page 4: ...o be safely drained away using barriers and drain trays if necessary 3 1 Water connection The connections for the various safety showers have a to 1 thread We recommend attaching the safety shower to...

Page 5: ...4 5 1 Models with safety eye showers combine the eye shower and body safety shower using the provided adapter 4 5 2 Models with hand held eye showers mount the bracket of the hand held eye shower at a...

Page 6: ...must be deinstalled during this process It must be reinstalled after the rinsing process The high performance shower head can be fastened and tightened using a pin type face wrench item no BR 010 100...

Page 7: ...the shower 9 3 Models with safety eye showers In order to start the eye shower press the big PUSH lever This locks the PUSH lever in the open position The eyewash should not make direct contact with t...

Page 8: ...is too low Water pressure is too low supply line cross section is too small or the jet regulator is contaminated The valve does not close properly The valve seal is damaged This is NOT a fault This is...

Page 9: ...eye shower irrespective of the existing operating pressure 2 2 Eye showers with two high performance spray heads To guarantee correct operation the flow pressure must be at least 2 bar The flow rate i...

Page 10: ...able fixing system 3 2 5 25 table mounted hand held eye shower Install using the 25 table mounting accessories M28x80mm A hole of 30 mm diameter should be made in the table top at a suitable point The...

Page 11: ...t possible to hold open the lids of the contaminated eyes with both hands at the same time The shower can also be removed from the bracket This allows other body parts or an injured person on the floo...

Page 12: ...have started to leak This is generally due to dirt from the water supply which collects in front of the valve If this has happened disconnect the handheld eye shower from the supply Operate the shower...

Reviews: