Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E
51
System Hookups
•
Verify the maximum load for any multiple portable socket-outlet (if used) does not
exceed 750 VA.
•
Do NOT place any multiple portable socket-outlets (if used) on the floor.
•
If multiple portable socket-outlets are used, DO NOT use the multiple socket-outlets
to power any equipment that isn't a part of the Welch Allyn OAE Hearing Screener.
(Adding other equipment may increase the current amount of leakage and exceed the
safety limit.)
•
Do not interconnect multiple pieces of equipment without verifying that the sum of all
leakage currents does not exceed the safety limit.
•
Do not connect any non-medical electrical equipment directly to a wall outlet if the
Welch Allyn OAE Hearing Screener is using a multiple, portable socket-outlet with a
separating iso-transformer. (The additional equipment may increase the current
amount of leakage and exceed the safety limit.)
•
Do not connect equipment which can potentially provide electromagnetic or other
types of interference. This may cause the Welch Allyn OAE Hearing Screener
equipment to function incorrectly.
•
Verify the equipment is connected to a power line source with the following frequency,
voltage and current capacity.
Frequency:
50 / 60 Hz
Voltage:
100 - 240 VAC
Current Capacity:
1.60 – 1.20A
WARNING!
Only use the Battery Charger Power Supply Models
Number 520-PS6VDC with the
Welch Allyn OAE Hearing Screener.
Summary of Contents for REF 29400
Page 1: ...OAE Hearing Screener Directions for Use REF 29400...
Page 14: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 13...
Page 16: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 15...
Page 17: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 16...