Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E
47
Electromagnetic Immunity
The REF 29400 Series OAE Hearing Screener is intended for use in the electromagnetic
environment specified below. The customer or user of the REF 29400 Series OAE Hearing
Screener should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-
4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines. IEC
61000-4-11
>95% dip in 0.5
cycle 60% dip in 5
cycles 30% dip for
25 cycles >95%
dip in 5 seconds
>95% dip in 0.5
cycle 60% dip in 5
cycles 30% dip for
25 cycles >95%
dip in 5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the REF 29400 Series OAE Hearing Screener
requires continued operation during power mains
interruptions, it is recommended that the REF
29400 Series OAE Hearing Screener be
powered from an uninterruptible power supply or
battery.
Power frequency
(50/60Hz)
magnetic field IEC
61000-4-8
3 A/m
40A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Summary of Contents for REF 29400
Page 1: ...OAE Hearing Screener Directions for Use REF 29400...
Page 14: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 13...
Page 16: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 15...
Page 17: ...Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E 16...