On Mach 5.0 PLUS only, remove the black plug from the air inlet on top of the
compressor module, and replace with the included filter.
CAUTION: Failure to replace the black plug with the filter will result in low air
condition, and damage to the compressor unit.
Air Supply Requirements (Mach 5.0 only)
For the Mach 5.0 only, be sure that the supplied air pressure and air flow volume to the
unit meet the minimum Air Requirements indicated in the Specifications section in
Chapter 1. Moisture filters must be used since compressed air lines may be
contaminated with moisture, oil or dirt, which are detrimental to the performance and
operation of the unit. Kreativ can supply a product (the Klean Pak) to assure a dry air
supply.
Caution: Connecting the instrument’s input air supply line to a water outlet will
lead to damage to the instrument which is not covered under the limited
warranty. Repairs for such damage are at the customer’s expense.
Electrical Requirements
Reference the Specifications section of Chapter 1 of this manual for the power
requirements of the unit. The power requirements of your instrument are located on the
system label on the back cover of the instrument. Verify the unit is set up correctly for
your electrical supply. When replacing fuses, only use fuses supplied by Kreativ in your
customer kit. If your requirements are different, consult Kreativ, Inc., or your dealer
before operating this instrument.
Mach 5.0/Dental Chair Pneumatic Integration
Please refer to Figure 3.
1. Depress chair foot switch while holding foot switch air line. The tube you feel
pulsing is the “foot switch return” line, mark it for later reference (step 5).
2. Close valve from office compressor to dental chair/unit and bleed pressure from
line.
3. Install “Tee 1” on line from compressor to dental unit before the regulator.
WARNING: The use of compressed oxygen as a gas supply for the unit
presents the possibility of having a gas that supports flammability. Under
NO circumstances should oxygen be used.