Page 17
990242 Rev. A
Footswitch Operation
•
Footswitch operation: depressing the footswitch starts the powder flow through the
handpiece/nozzle.
Unit Shutdown
•
Set the Power Switch to OFF.
Unloading Powder Chamber
The illustrations on page 16 show the parts involved in unloading the Mixing Chamber.
Follow these directions when unloading it.
•
Open the back access panel.
•
Firmly hold the chamber and vibrator and loosen the Mixing Chamber Lid by turning it
counterclockwise.
•
Completely unscrew the lid from the Mixing Chamber.
•
Loosen the locking knob on the Chamber Tilt Plate; slide the Chamber Clamp back to
release and tilt the chamber out of the cabinet.
NOTE:
A container must be placed under the chamber opening to catch the abrasive powder.
•
Set the Beam Intensity on the front panel to “10”.
•
Press the Mixing Chamber Unloading Switch to vibrate the abrasive powder out of the
Mixing Chamber.
•
Tilt the Mixing Chamber back to its upright position.
•
Slide the Chamber Clamp forward over the Chamber Tilt Plate and tighten the knob.
•
Replace the lid and nut making sure the threads are free of powder and the lid gasket
is properly seated.
•
Check the two air hoses attached to the Mixing Chamber to assure that they are not
kinked or pinched and that they make a large free loop.