Press the
Key and the display will stop flashing.
Use the dB HL dial to adjust the Probe Tone Level to 85.5 dB SPL (+/- 0.9
dB). At this point the Probe Tone Frequency may be verified. It should equal
226 Hz (+/- 6 Hz).
5)
Press the
Key to store the data in memory. The display will update to
the following ‘Washing” display.
At this point place the probe in the 2 cc test cavity.
6)
Press the
Key to start the microphone calibration process. The
display
will update to the following steady state display.
After a few seconds the SAVED indicator will extinguish and the 1.2 indicator
will change to 1.3. When the cycle is complete the display
will update to the
following “flashing” display.
7 )
Press the
Key to return to the Main Calibration Mode Menu.
COMPLIANCE CAL
1)
Use the
Hz
to position the cursor
Mode position.
2)
Press the
Key to enter the Compliance
update to the following “flashing” display.
at the Compliance Cal
Cal Model The display will
Summary of Contents for GSI 38
Page 1: ......
Page 3: ...1 t t L c c k 4 c i I e PRODUCT SPECIFICATION 2...
Page 14: ...i i L i i c FUNCTIONAL DESCRIPTION 13 c...
Page 29: ...CIRCUIT THEORY L 28...
Page 36: ...i L c i k t CALIBRATION t 35...
Page 51: ...TROUBLE SHOOTING L b L I 4 50...
Page 57: ...L L L k DISASSEMBLY 56...
Page 59: ......
Page 60: ......
Page 63: ...I L L c t t AUDIO TYMP BOARD ASSEMBLY DRAWINGS SCHEMATICS 62...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 80: ...PROBE 79...
Page 81: ......
Page 82: ......
Page 83: ......
Page 87: ...L L L L PUMP 86...
Page 88: ......
Page 89: ...i PRINTER e 88...
Page 90: ......
Page 92: ...I I c DISPLAY L L i i 91...
Page 93: ......
Page 95: ...I I c ASSEMBLY L L i i 94...