Code Number
09
10
12
‘13
'14
'15
‘16
‘17
'18
'19
‘20
‘21
22
23 Calibration Data Write Error occurred
24 Invalid Present Bar Operation due to a
Calibration Data Error
Description/Probable Cause
Conf. Register Error if non-recoverable re-calibrate Config,
Register then replace Microprocessor U34.
Halt in Response to Remote Command / if non-recoverable
possible defective Audio/Tymp Communications Chip U36
or Microprocessor U34.
Invalid Keycode detected in CMDTSK / if non-recoverable
possible defective Switch Panel, Switch Buffer U26, U27, or
U28, or Microprocessor U34.
Invalid Keysource detected in CMDTSK /
same as above.
Invalid
above.
Queue Power
Error detected in CMDTSK /same as
Bad Pointer to EEPROM Routine’/
Collision of the Normal
Stack and Swap
Stack /
Invalid function for
a Banded Swapped Function
/
Printer
RAM
Error on power up / internal or micro
Tymp Hardware State is unknown /
Non-Fatal Audiometer related error messages
Calibration Data Read Error occured /
HL Limit
exceeded /
Invalid Extended Range Selection due to HL not in range
where the Extend Range is valid /
Summary of Contents for GSI 38
Page 1: ......
Page 3: ...1 t t L c c k 4 c i I e PRODUCT SPECIFICATION 2...
Page 14: ...i i L i i c FUNCTIONAL DESCRIPTION 13 c...
Page 29: ...CIRCUIT THEORY L 28...
Page 36: ...i L c i k t CALIBRATION t 35...
Page 51: ...TROUBLE SHOOTING L b L I 4 50...
Page 57: ...L L L k DISASSEMBLY 56...
Page 59: ......
Page 60: ......
Page 63: ...I L L c t t AUDIO TYMP BOARD ASSEMBLY DRAWINGS SCHEMATICS 62...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 80: ...PROBE 79...
Page 81: ......
Page 82: ......
Page 83: ......
Page 87: ...L L L L PUMP 86...
Page 88: ......
Page 89: ...i PRINTER e 88...
Page 90: ......
Page 92: ...I I c DISPLAY L L i i 91...
Page 93: ......
Page 95: ...I I c ASSEMBLY L L i i 94...