ELI 150
6-16
Set the “Default XMT Media” to WLAN and install your network information; the following settings are required
in the cardiograph and should be obtained from your local network personnel:
Host IP address
(network specific)
Port #:
5005
SSID:
(access point name)
Channel Number: 6
WEP info (if used)
Once this setup is completed ECGs can be transmitted by pressing the “XMT” key from the main menu and
selecting the record you wish to send.
Transmission information can be viewed in the test software window.
When completed the UUT display will return to the ECG directory and a “T” flag will be set indicating proper
transmission. The transmitted ECG can be seen in the target directory (C:\UNIPRO) listed as a new file.
Safety Tests
If the cardiograph housing was opened for repair or inspection work, the following safety tests should be performed in
accordance with the EN/IEC 60601-1 or EN 62353/IEC 62353 methods and limits.
In countries with different national legislation, the country specific standards and guidelines for retesting of electrical
medical devices are obligatory.
Objective evidence of testing results versus the defined standards should be provided as part of the servicing record.
The following tests are performed per the above specifications:
AC Hi-Pot Patient to Mains
Earth Leakage
Enclosure Leakage
Patient Source
Patient Sink
Summary of Contents for ELI 150
Page 3: ......
Page 10: ...viii...
Page 14: ...xii...
Page 36: ......
Page 44: ...ELI 150 4 8 Re assemble in reverse order...
Page 54: ...ELI 150 4 18 Re assemble in reverse order...
Page 72: ......
Page 74: ...ELI 150 5 2 ELI 150 Block Diagram...
Page 77: ...Section 5 5 5 Specifications...
Page 80: ......
Page 100: ......