Remove the Philips screw, item # 7, from the print head assembly.
Remove the print head from the print head mount.
Remove the cables, item # 35 (qty 2) from the print head.
Note:
An Antistatic brush is attached to the print head, item # 34. This brush must be installed on print head prior to
reassembly. Failure to install the anti-static brush can cause premature failure of the print head.
Re-assemble in reverse order.