Protection against electrosurgical
Electrosurgical protection per EN60601-2-27:2011. ECG device will resume as per the
table below.
Parameter
Initial report after
defibrillation
Maximum time to
recover
Note
HR value
Acquiring
10 sec +5 beats
Depends on patient heart
rate
ECG waveform
Waveform
≤
10 sec
Respiration value
Acquiring
10 sec +5 breaths
Depends on patient
respiration rate
Pacer detect
Pacer marker in waveform
≤
10 sec
Arrhythmia
Acquiring or
Cannot analyze
10 sec +5 beats
Depends on patient heart
rate
Device classification
ECG device will operate per table below after startup and after changes to parameter
configurations (e.g., filter on/off) or lead off recovery.
Parameter
Initial report after
defibrillation
Maximum time to
recover
Note
HR value
Acquiring
10 sec +5 beats
Depends on patient heart
rate
ECG waveform
Waveform
≤
3 sec
Respiration value
Acquiring
10 sec +5 breaths
Depends on patient
respiration rate
Pacer detect
Pacer marker in waveform
≤
3 sec
Arrhythmia
Acquiring or
Cannot analyze
10 sec +5 beats
Depends on patient heart
rate
Default settings
Characteristic
Specification
Lead color code
AHA
Lead for view
II
Sweep speed
25 mm/second
Directions for use
Specifications 33
Summary of Contents for Connex ECG Module
Page 4: ...iv Contents Welch Allyn Connex Devices ECG Module...
Page 10: ...6 Screen elements Welch Allyn Connex Devices ECG Module...
Page 22: ...18 Alarms Welch Allyn Connex Devices ECG Module...
Page 32: ...28 EMC guidance and manufacturer s declarations Welch Allyn Connex Devices ECG Module...
Page 41: ......