Characteristic
Specification
Noise suppression – DC current of 200nA max for RL
Software and alarm specifications
Characteristic
Specification
Ventricular tachycardia threshold default
120 beats per minute ±3 beats per minute for 6 consecutive beats
Variable ventricular tachycardia settings
100 beats per minute to 150 beats per minute ±3 beats per minute for 6
consecutive beats
Asystole threshold default
No detectable beat for
≥
4 seconds
Ventricular fibrillation threshold default
A fibrillatory waveform persisting for
≥
4 seconds
Configurable mains filter
ON or OFF; 50 Hz or 60 Hz
Data stream bytes output
Maximum latency <100 ms
Heart rate averaging method per IEC
60601-2-27:2011 Clause 201.7.9.2.9.101
Number of consecutively detected beats:
>10: Heart rate = 60/(average of last 10 detected consecutive R-R
intervals)
>2 and <10 : Heart rate = 60/(average of the detected consecutive R-R
intervals)
Response to change in Heart rate per IEC
60601-2- 27:2011 Clause 201.7.9.2.9.101
Increasing from 80 to 120 beats/minute: 6.0 sec
Decreasing from 80 to 40 beats/minute: 16.0 sec
Protection against defibrillation
Defibrillation protection per EN60601-2-27:2011. ECG device will resume according to
the table below.
Parameter
Initial report after
defibrillation
Maximum time to
recover
Note
HR value
Acquiring
3 sec +5 beats
Depends on patient heart
rate
ECG waveform
Waveform
≤
5 sec
Respiration value
Acquiring
3 sec +5 breaths
Depends on patient
respiration rate
Pacer detect
Pacer marker in waveform
≤
5 sec
Arrhythmia
Acquiring or
Cannot analyze
5 sec +5 beats
Depends on patient heart
rate
32 Specifications
Welch Allyn Connex
®
Devices ECG Module
Summary of Contents for Connex ECG Module
Page 4: ...iv Contents Welch Allyn Connex Devices ECG Module...
Page 10: ...6 Screen elements Welch Allyn Connex Devices ECG Module...
Page 22: ...18 Alarms Welch Allyn Connex Devices ECG Module...
Page 32: ...28 EMC guidance and manufacturer s declarations Welch Allyn Connex Devices ECG Module...
Page 41: ......