Characteristic
Specification
Mains filter
50 Hz, 60 Hz, off (default = 60 Hz)
Frequency range
0.5 to 70 Hz
Power source
USB (4.5 V to 5.5 V)
Digital data interface
Serial (USB – Full Speed)
Monitoring period
Continuous
ECG channels
Supports 3-lead or 5-lead
Sweep speed
25 mm/sec, 50 mm/sec
Resolution
14 bit ECG data, 30 mV peak-to-peak dynamic range minimum
Lead-off detection
3-lead or 5-lead
Applied currents
Lead connection failures are detected by a bias current with respect to
the reference electrode.
User-adjustable parameters
See Alarms
ECG cable specifications
Characteristic
Specification
Connector at ECG module
80 in. ±1 shielded cable to a 24 in. ±1 lead fan-out to electrode
connection
ECG cables
3-lead or 5-lead to snap connection
Patient cables
Conforms to ANSI/AAMI EC53
Impedance respiration specifications
Characteristic
Specification
Impedance Respiration accuracy
±2 breaths per minute or ±2%, whichever is greater
Range
5 to 100 breaths per minute
Breath detection range
0.4 ohms to 3.0 ohms
Respiration lead source
Lead II (right arm and left leg)
Note
Respiration can only be acquired from Lead II.
Respiration, leads-off sensing and active
noise suppression (applied currents)
Respiration – Excitation signal characteristics Less than 25 uA RMS at
31 kHz pseudo sine wave
Lead off – DC current of 50nA max for RA ,LA, LL, V; 200nA max for RL
Directions for use
Specifications 31
Summary of Contents for Connex ECG Module
Page 4: ...iv Contents Welch Allyn Connex Devices ECG Module...
Page 10: ...6 Screen elements Welch Allyn Connex Devices ECG Module...
Page 22: ...18 Alarms Welch Allyn Connex Devices ECG Module...
Page 32: ...28 EMC guidance and manufacturer s declarations Welch Allyn Connex Devices ECG Module...
Page 41: ......