8
Test the device
4. Test the device
Important!
Perform the following test on the device in accordance with test instruction WM 22401 after every repair and on
servicing. The device test can also be used for troubleshooting.
If you find faults or deviations from specified values during testing, you may not use MODUL CapnoVol until
the faults have been rectified.
We recommend always keeping the following in stock:
• replacement seals for device connections WM 22468, WM 22469
• Ni-Mh battery WM 22458
4.1 Test material required
• Ventilation device, e.g. MEDUMAT with the associated test set, e.g. WM 15382
• KAL Check gas WM 97061 with pressure reducer WM 97062
• PC with installed software WM 22480 from the following set WM 15681
• Set of IR adapter plus software and bracket WM 15681
• Power supply cable, 12 V, WM 22895
• Connecting cable for flow measurement WM 22420
• Measuring and leaktightness adapter WM 22868
• Current measuring adapter MODUL CapnoVol WM 22817
• CO
2
absorber WM 22172
• Set for testing power supply MEDUMAT/MODULES WM 15440
• Set of tools for MEDUMAT Standard WM 15349 or syringe set WM 15359
• Volume flow measuring device (e.g. PF-300), breath volume measurement with O
2
in STP mode
• Digital multimeter
• Patient tube with patient valve
• Pall filter
4.2 Preparation for testing
1. Fit MODUL CapnoVol together with a
MEDUMAT ventilation device to a LIFE-BASE
carrying system.
2. Connect the external power supply.
3. Connect sensor cable
.
4. Connect 12 V power supply cable
Summary of Contents for Emergency MODUL CapnoVol