Repair information and repair instructions
23
4. Take the display illumination cable connector
off the main circuit board.
5. Take connector
X8
off the main circuit board.
6. Undo the ribbon cable from latch
X9
: to do so,
pull up the top part of the latch. You can then
pull out the cable.
Warning!
Ensure that the main circuit board and the
cable are not damaged in the process.
7.3 Close the device
Materials and tools required
• ESD workstation
• Phillips screwdriver, size 2
1. Push the ribbon cable into latch
X9
on main
circuit board
and then push the top part of
the latch back in.
2. Push connector
X8
onto the main circuit board
3. Push the display illumination connector onto
main circuit board
4. Position the insulating cover back inside the
device over the main circuit board and CAP
100 circuit board and glue it firmly in place
with Loctite WM 14920.
5. Carefully put top part of housing
6. Turn the device over and screw the 4
into the bottom part of the housing.
7. Test the device (see “4. Test the device” on
X8