22
Repair information and repair instructions
7. Repair information and repair instructions
7.1 General
Only perform repairs to MODUL CapnoVol at an ESD workstation!
•
Follow the safety instructions in the instructions
for use for MODUL CapnoVol.
• Any handling of this device assumes in-depth
knowledge and observance of the instructions for
use and of the servicing and repair
instructions.
• Only perform repairs described in these
servicing and repair instructions. This is the
only way to guarantee that
MODUL CapnoVol works perfectly.
• Always perform a test after any repair or
servicing work (see “4. Test the device” on
page 8).
• If you replace components or individual parts,
use only genuine
WEINMANN replacement
parts
.
•
Note
The item numbers listed in the text which
follows are identical to the item numbers in the
replacement partslist on Page 35 and in the
overview on Page 4.
7.2 Open the device
Materials and tools required
• ESD workstation
• Phillips screwdriver, size 2
• Pointed pliers
1. Place the device on a non-slip surface and
unscrew the 4 screws
.
2. Carefully pull off top part of the housing
and swing it away.
3. Pull the insulating cover out of the inside of the
device.
Note
The insulating cover is glued. If the insulating
cover (WM 22387) is damaged when pulled
out, it must be replaced before the device is
closed.
Insulating
cover
Summary of Contents for Emergency MODUL CapnoVol
Page 43: ......