5.4 Change O-rings of oxygen duct inlet
Materials and tools required
• Watchmaker's screwdriver (cleaned with
spirit)
• Allen key, 5 mm
1. Unscrew threaded connection
from oxygen
device using the Allen key.
3. Lift the O-rings
carefully out of the two
connections using the cleaned watchmaker's
screwdriver.
Be careful not to damage the connections.
4. Clean the connections using a dry cloth or a
cloth moistened with clean water.
5. Push the new O-rings carefully into the
connections in question. Do not use a tool for
this.
6. Screw sealing screw
connection on the rear of the device.
7. Use the Allen key to screw threaded
onto oxygen duct inlet
5.5 Calibrate O
2
sensor
5.5.1 Perform calibration
Materials and tools required
• Service set for MODUL CapnoVol
WM 15743 consisting of:
– level converter WM 22825
– calibration software WM 22919
– CAP 100 external test connecting cable
WM 22855
– Plug-in power supply unit WM 22859
• CO
2
absorber WM 22172
• CO
2
gases
– 2.5 % CO
2
± 0.2 % in O
2
– 5 % CO
2
± 0.2 % in O
2
– 9.7 % CO
2
± 0.2 % in O
2