No
�
Fault
Can be r
ecogniz
ed b
y
Possible cause
Remedial measur
es
3
Segments of the clamping jaws do not separate widely enough
Connector is seated too loosely
, it sits unsteadily
in the test piece
W
ear
Check
test piece for damage
If it
is
in
good
condition, send the
connector to
WEH for maintenance
Connector is not corr
ect
-
ly connected
Repeat the connecting pr
ocedur
e
4
Connector wobbles despite a corr
ect connection to the
test piece
Incorr
ect connection
geometr
y at test piece
W
ear on geometr
y of the
test piece
’s connection
par
ts
Replace test piece
W
ear on the connector
’s
clamping jaws
Return the connector to WEH
for maintenance
5
Clamping jaws deformed
The connector can not be plugged into the test piece
Ex
cessiv
e mechanical
str
ess due to inappr
opri
-
ate handling
Return the connector to WEH
for maintenance
If y
ou should encounter any other pr
oblems, please contact WEH or y
our r
esponsible distributor
.
MD-10320-L51-R1.0.0-03
Page 27
Operating instructions
AE