8�2
Connection
Î
Apply pilot pressure to the connector. The clamping jaws
will close together and the connection to the test piece is
possible (
Î
Grip the test piece and plug it up to the stop straight onto
the connector (
Doing so will prevent any damage to the front seal caused
by any sharp edges on the test piece and ensure a correct
connection.
Attention:
Insert the test piece carefully and not with force
onto the connector.
Î
Vent the pilot pressure line. The test piece tightens itself to
the connector and creates a pressure-tight connection.
Attention:
Make sure that the clamping jaws on the test
piece grip correctly, that the connector is seated prop-
erly (the test piece must not wobble) and that the pres-
sure-tight connection has been made. Only then is the
connector connected correctly. Otherwise the test piece
may spring out of place and medium may escape.
Î
Pull lightly on the test piece (red arrow
)
in the direction opposite that of the connection to make
sure that the connector is firmly in place.
- The connector is now connected pressure-tight to the test
piece.
Î
To start the testing procedure, apply pressure to media
inlet ‘B1’.
Figure 3
Figure 2
Figure 4
pull backwards
MD-10320-L51-R1.0.0-03
Page 17
Operating instructions
AE