A35SE1-2
WEBECO Hygiene in Medizin und Labor GmbH und Co. KG
51
Vers. 02 / 04.98
11. Operation Quick Guide
1.
The water container has to be filled with demineralized
or destilled water. Switch on mains.
2.
Load sterilizer, (see chapter 7.2.2.1).
Please see chapter 7.3 for loading instructions.
3.
Close door and firmly turn closing spindle clockwise.
4.
Turn the red drag pointer of the pressure gauge (1/1)
left until „0“ is displayed.
5.
Start suitable program according to kind of load to be
sterilized by pressing the program-start-key (2/1;2/2).
The running sterilization-cycle is shown through the
program-display (2/3;2/4) above the „Start“ key.
6.
After the start of the sterilization-cycle the residual-time-
display (2/5) shows the remaining time until the
„Program End“.
7.
The end of the program is indicated by a blinking
program-display (2/3;2/4).
The sterilization material may be unloaded.
While pressure reading >0< (black pointer) it has to be
tested, if the necessary sterilization pressure 1,1 bars -
121°C, 2,1 bars - 134°C, has been reached.
Otherwise the sterilization load must always be
considered non-sterile.
8.
Turn the closing spindle counter clockwise for opening
of the door (see chapter 7.5).