WEBECO Hygiene in Medizin und Labor GmbH und Co. KG
A35SE1-2
Vers. 02 / 04.98
36
7.4 Notification- and Failure Reports
Functions and operating conditions of the sterilizer are
continuously controlled by the internal computer control.
In case of malfunctions or deviations an alarmcode is
displayed in the residual time display.
7.4.1 Indications
Text in
Display
Cause
Countermeasure
s
A1
pressure switch
defective
cable break
pressure switch
defective
Notify customer
service
A2
Pressure Increase too
long
Water Shortage
Untightness
Heating defect
Refill Water
Notify customer
service
Notify customer
service
A3
Exceeding reswitching
time of Heating
Element
Water Shortage
unsuitable
Loading
Too much
Loading
Refill Water
see Operating
Instruction chapter
7.2.2
Reduce Loading
A4
Failure Power Supply
Failure of Power
Supply
Test Power Supply
existing supply
voltage notify
customer service
ATTENTION!
If failures are indicated and the cause can not be rectified,
the appropriate customer service is to be informed.
CAUTION!
If a program ends with a failure, the sterilization material is
considered to be non-sterile.