A35SE1-2
WEBECO Hygiene in Medizin und Labor GmbH und Co. KG
39
Vers. 02 / 04.98
7.5 Automatic Safety Lock
To ensure the safety of the operator, the sterilizer chamber
is automatically secured by a door locking mechanism. The
condition of this mechanism is indicated by the yellow
display (2/8).
7.5.1 Door Spindle Interlock
At the start of a program cycle (pushing the „Start“ button(2/1;
2/2)), the door spindle is locked automatically. In this
condition, the door spindle may only be turned one quarter
rotation counter clockwise, until the mechanism locks.
The door spindle is not allowed to be tightened up or further to
be closed during the running program.
7.5.2 Unlocking of Door Spindle
When switching on the sterilizer or at end of one program the
door interlock device system is released and the door may be
opened. This condition is indicated by the shining of the yellow
display (2/8).
INDICATION:
If an attempt to open the door has been made while a
sterilization process was in progress, please take note of the
following:
1.
After correct program run „Program End“ release the
locking device mechanism by applying a slight turn.
2.
The door locking device system can then be opened as
usual, by turning the spindle counter clockwise.