A35SE1-2
WEBECO Hygiene in Medizin und Labor GmbH und Co. KG
41
Vers. 02 / 04.98
7.7 Activating/Deactivating Additional Cooling
The setup-procedure informs the apparatus if cooling-water
is connected as a supplemental cooling or not.
At every change of the method of operation the setup
procedure has to be carried through.
The set-up procedure is initialized if the left key (2/1) is
pressed when switching on the main switch (1/2). As
confirmation „c“ is shown in the display of the residual time.
After releasing the key „00“ or „01“ is shown.
The meaning of the numbers is as follows:
00
:
additional cooling off
01
:
additional cooling on
When using the right key it is possible to switch between „0“
and „1“. Additional using of the left key (2/1) confirms the
choice.
After releasing of the key the apparatus is in its normal
operating condition. The A 35-S can work in continuos
operation with or without supplemental cooling. Thus the
process-time is shortened by approx. 10-15%.
When working without supplemental cooling after 3
sterilizations directly after one another a reduced drying
capacity or steam disturbance from the water reservoir is
possible.
IMPORTANT:
Operation with supplemental cooling without water supply
leads to highly reduced drying capacity.